Cardio Wave Analyzer Arterial and Stress Testing Device

CardioWA Specification

CardioWA Specification

Specifications for the CardioWA

The CardioWA comes with the following:

A Finger Probe

USB Interface to a computer

Latest Max Pulse Software 

Free shipping, handling and insurance in the continental US

A one year warranty

  • Specifications for the Cardio Wave Analyzer (CardioWA)
  • Size: Width: 6 inches  Depth: 5.5 inches Height: 1.5 inches)
  • Weight:  1.43 pounds
    Technical Specifications
  1. PPG (photoplethysmography)
    Measurement Range 30-240 BPM
    Accuracy +/- 2 %
    Wave-out time 2 seconds
    Averaging (after setting time) 8 beats
    Pulse Wave / Heart Rate
  2. Display Parameters
    – Pulse Wave
    – Accelerated Pulse Wave
    -Condition of the blood vessels
    -HRV (Heart Rate Variability)
  3. Environmental
    Temperature Range
    -Operating 10 – 40 degrees C (50 to 104 degrees F)
    -Storage -20 – 60 degrees C (-4 to 140 degrees F)
    Relative Humidity Range
    -Operating 80% relative humidity max @ 40 degrees C (104 degrees F) (without condensation)
    -Storage 95% relative humidity max @ 50 degrees C (122 degrees F)
    Atmospheric Pressure Range
    -Operating 700 -1060 hPa (hectoPascals)
    Electromagnetic Compatibility
    -Radiated and conducted electromagnetic energy per CISPR 11, Class A
  • Standards
  1. Safety
    -Protection class 1 (The basic means of protection is the insulation between live parts and exposed conductive parts such as the metal enclosure)
    -Degree of Protection Type BF (generally  for devices that have  conductive contact with the patient)
  2. Electrical
    -Power Requirements: 100 – 240 VAC, 50/60 Hz, 1.0 A
    -Power Fail Protection
  • Warranty
  • 1 Year from date of purchase.